Product Data Sheet

Model 588B 12 or 24-Channel Rackmount Pulse Generator

A rackmount pulse generator built for high channel density. Twelve channels in a 1U chassis or twenty-four in a 2U chassis, with up to thirty-six independent outputs, 250 ps delay resolution, and ultra-low jitter below 5 ps RMS for precision multi-channel synchronization.

Berkeley Nucleonics Corporation
Berkeley Nucleonics Model 588B 24-channel pulse generator, front panel

1Overview

Verification notes. Values are drawn verbatim from the Model 588B datasheet (V1.1, dc-11605202). Confirm against the current released datasheet before quoting. Output-module figures depend on the modules ordered.

The Model 588B Pulse Generator has a powerful set of functions that serve various modes of operations and applications. Both the system timer and the channel timer combine to provide the ability to generate complex waveforms. BNC offers the 1U unit for 12 channels or the 2U unit for 24 channels.

These units provide precision timing and synchronization to effectively trigger any series of events or necessary equipment with up to 36 independent outputs, dual inputs with gate/trigger or optional dual trigger, and an external clock reference input. Delay resolution is 250 ps, delay accuracy is 1 ns plus 0.0001 times the period, and channel-to-channel jitter stays below 5 ps RMS.

2Key Features

  • 250 ps delay resolution
  • < 5 ps RMS jitter
  • 1 ns + .0001 x period delay accuracy
  • USB / RS232 / Ethernet communications
  • Twelve channels (1U) or twenty-four channels (2U), up to thirty-six independent outputs
  • System timer plus per-channel timers for complex waveforms
  • Dual inputs with gate/trigger or optional dual trigger
  • External clock reference input

3Channels & Timing Architecture

The 588B is built around a system timer with independent per-channel timers. Each channel sets its own delay and width, so a single instrument can drive a dozen or two dozen precisely timed edges across one experiment. The case for high density is straightforward: where a four-channel or eight-channel instrument would need three units to reach twenty-four edges, the 588B does it in one 2U chassis, and every channel shares the same low-jitter time base.

Delay covers 0 to 2,000 s and width covers 10 ns to 2,000 s, both at 250 ps resolution. An external clock reference input lets the instrument lock to a system clock, keeping all channels coherent with the rest of the setup. The ultra-low jitter below 5 ps RMS is what makes the high channel count usable for precision synchronization rather than just count.

Model 588B rear panel
Rear panel. High-density channel outputs, trigger and gate inputs, and clock reference.

4Specifications

Internal Rate Generator

ParameterValue
Rate (T0 period)50 ns to 5,000 s (0.0002 Hz to 20.000 MHz)
Resolution5 ns
Accuracy1 ns + .0001 x period
Jitter< 5 ps RMS
Settling1 period
Burst Mode Counter1 to 4,000,000,000 pulses
Duty Cycle Mode Counters1 to 4,000,000,000 pulses
Cycle Counter1 to 10,000,000 cycles
Timebase200 MHz, low jitter PLL
Oscillator50 MHz, 25 ppm
System Output ModesSingle pulse, burst, duty cycle, external gate/trigger
Pulse Contol ModesInternal rate generator, external trigger/gate

Channel Timing Generator

ParameterValue
Channel Output ModesSingle shot, burst, duty cycle, normal
Control ModesInternally triggered, externally triggered and external gate. Each channel may be independently set to any of the modes.
Burst Mode Counter1 to 10,000,000 pulses
Duty Cycle Mode Counters1 to 10,000,000 pulses
Wait Function Counter0 to 10,000,000 pulses
Output MultiplexerUp to five (5) channel timers may be routed to each output channel.
TimebaseSame as internal rate generator

Delays

ParameterValue
Delay Range0 - 2,000 s
Width Range10 ns - 2,000 s
Accuracy1 ns + 0.0001 x Delay
Resolution250 ps
Pulse Inhibit Delay< 120 ns typical
Output Inhibit Delay< 50 ns typical

5Output Module Specifications

The standard output is the TTL/Adjustable Dual Channel Output Module, with an output impedance of 50 ohm. It operates in either a fixed TTL/CMOS mode or an adjustable mode.

TTL/CMOS Mode

ParameterValue
Output Level4.0 V typ into 1 kohm
Rise Time3 ns typ (10% - 90%)
Slew Rate> 0.5 V/ns
Jitter50 ps RMS channel to channel

Adjustable Mode

ParameterValue
Output Level2.0 to 20 VDC into 1 k ohm; 1.0 to 10.0 VDC into 50 ohm
Output Resolution10 mV
Current200 mA typical, 400 mA (short pulses)
Rise Time15 ns typ @ 20 V (high imp); 25 ns typ @ 10 V (50 ohms) (10% - 90%)
Slew Rate> 0.1 V/ns
Overshoot< 100 mV + 10% of pulse amplitude

6Triggering & Inputs

System External Trigger/Gate Input(s)

Trigger Input

ParameterValue
TypePulse 0.5 to 30v, switch closure (footswitch, safety interlock, etc.)
FunctionGenerate individual pulses, start a burst or continuous
RateDC to 1/ (200 ns + longest active pulse). Maximum of 5 MHz
SlopeRising or Falling

Gate Input

ParameterValue
ModePulse inhibit or output inhibit
PolarityActive high/active low
Channel BehaviorGlobal w/Individual Channel Control

Trigger/Gate Dual Input Module (Standard)

Standard dual channel input module, providing one trigger input, one gate input and one dual purpose input on the front panel.

ParameterValue
Threshold0.2 to 15 VDC
Maximum Input Voltage30 V Peak
Impedance5.3 K ohm + 10pF
Resolution10 mV

Trigger Input

ParameterValue
SlopeRising or Falling
Jitter800 ps RMS
Insertion Delay< 160 ns
Minimum Pulse Width10 ns

Gate Input

ParameterValue
Pulse Inhibit Delay< 160 ns
Output Inhibit Delay< 160 ns

7Standard Features & General

Standard Features

ParameterValue
CommunicationsUSB / RS232 / Ethernet
External Clock In10 MHz, 20 MHz, 25 MHz, 30 MHz, 40 MHz, 50 MHz, 60 MHz, 80 MHz
External Clock OutT0, 10 MHz, 20 MHz, 25 MHz, 30 MHz, 40 MHz, 50 MHz, 60 MHz, 80 MHz, 100 MHz

General

ParameterValue
Storage12 ch = 12 storage bins; 24 ch = 24 storage bins
Dimensions19" x 10" x 3.50" 12/24 ch (standard 2U rack mount)
Weight8 lbs
Power100 - 240 VAC, 50/60 Hz, < 3 A
Fuse(Qty 2) 630 mA, 250 V Time-lag

8Options & Configurations

  • Twelve channels (1U) or twenty-four channels (2U), up to thirty-six independent outputs
  • Dual inputs: gate/trigger or optional dual trigger
  • TTL/Adjustable Dual Channel Output Module (standard)

The full options and ordering chart is not published on the product page and should be confirmed with Berkeley Nucleonics when configuring a unit (verify).

9Applications

  • ICCD and PIV testing
  • Laser triggering and gating
  • Pulsing devices under test and pump lasers
  • Radar and sonar simulation
  • High-speed photography

10Ordering / In the Box

Specify channel count (twelve or twenty-four), output-module type, and input options when ordering. The instrument ships as a 19-inch rackmount unit with power cord and a calibration certificate.

ItemDetail
Model 588B, 12-channelTwelve independent channels, 1U chassis, 250 ps resolution
Model 588B, 24-channelTwenty-four independent channels, 2U chassis, up to 36 outputs

To configure or quote, contact Berkeley Nucleonics at info@berkeleynucleonics.com or 800-234-7858.